Journal: Journal of Medical Imaging
Article Title: Comparison of the CTDI and AAPM report No. 111 methodology in adult, adolescent, and child head phantoms for MSCT and dental CBCT scanners
doi: 10.1117/1.JMI.4.3.031212
Figure Lengend Snippet: Summary of the three parameters D eq , L eq , and α from the AAPM fit in (a) Toshiba MSCT scanner and (b) i-CAT Next Generation dental CBCT scanner. (a) For each patient group, the D eq values from thimble chamber measurements are very close to each other, with the highest being present in the child head phantom and the lowest in the adult head phantom. The pencil chamber measurements generate larger D eq , α , and L eq values. (b) The AAPM method and the two Sedentex methods result in similar D eq , α , and L eq values for each patient group.
Article Snippet: This study has examined three CT imaging scanners, including one common clinical MSCT scanner (Toshiba Aquilion TM One, Toshiba America Medical Systems, Inc., Tustin, California) and two dental CBCT scanners (i-CAT Next Generation CBCT, Imaging Sciences International, LLC, Hatfield, Pennsylvania and CS 9300 CBCT, Carestream Dental LLC, Atlanta, Georgia).